Patellar Tendon Repair Augmentation With a Knotless Suture Anchor Internal Brace: A Biomechanical Cadaveric Study

Am J Sports Med. 2018 Apr;46(5):1199-1204. doi: 10.1177/0363546517751916. Epub 2018 Feb 5.

Abstract

Background: Patellar tendon repair with braided polyethylene suture alone is subject to knot slippage and failure. Several techniques to augment the primary repair have been described. Purpose/Hypothesis: The purpose was to evaluate a novel patellar tendon repair technique augmented with a knotless suture anchor internal brace with suture tape (SAIB). The hypothesis was that this technique would be biomechanically superior to a nonaugmented repair and equivalent to a standard augmentation with an 18-gauge steel wire.

Study design: Controlled laboratory study.

Methods: Midsubstance patellar tendon tears were created in 32 human cadaveric knees. Two comparison groups were created. Group 1 compared #2 supersuture repair without augmentation to #2 supersuture repair with SAIB augmentation. Group 2 compared #2 supersuture repair with an 18-gauge stainless steel cerclage wire augmentation to #2 supersuture repair with SAIB augmentation. The specimens were potted and biomechanically loaded on a materials testing machine. Yield load, maximum load, mode of failure, plastic displacement, elastic displacement, and total displacement were calculated for each sample. Standard statistical analysis was performed.

Results: There was a statistically significant increase in the mean ± SD yield load and maximum load in the SAIB augmentation group compared with supersuture alone (mean yield load: 646 ± 202 N vs 229 ± 60 N; mean maximum load: 868 ± 162 N vs 365 ± 54 N; P < .001). Group 2 showed no statistically significant differences between the augmented repairs (mean yield load: 495 ± 213 N vs 566 ± 172 N; P = .476; mean maximum load: 737 ± 210 N vs 697 ± 130 N; P = .721).

Conclusion: Patellar tendon repair augmented with SAIB is biomechanically superior to repair without augmentation and is equivalent to repair with augmentation with an 18-gauge stainless steel cerclage wire.

Clinical relevance: This novel patellar tendon repair augmentation is equivalent to standard 18-gauge wire augmentation at time zero. It does not require a second surgery for removal, and it is biomechanically superior to primary repair alone.

Keywords: internal brace; knee; knotless repair; ligament; patellar tendon; tendon.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Biomechanical Phenomena
  • Cadaver
  • Female
  • Humans
  • Male
  • Middle Aged
  • Patella / surgery
  • Patellar Ligament / injuries*
  • Patellar Ligament / surgery*
  • Suture Anchors*
  • Suture Techniques*
  • Sutures
  • Tendon Injuries / surgery*